Ozone In Air

Ozone in air measurement

Ozone in Air Analyser: Eco Sensors UV-100

Measuring ozone in air means determining the concentration of ozone gas (O3) in a sample of air, usually in parts per million (ppm) or parts per billion (ppb) by volume. The most accurate way to do it is UV photometry: ozone absorbs ultraviolet light at 254 nm, so the amount of light absorbed tells you exactly how much ozone is present.

The Eco Sensors UV-100 is a compact UV photometric ozone analyser that reads from 0.01 to 999 ppm with 2% accuracy. It logs data, drives 4–20 mA and 0–2.5 V outputs, and warns you when ozone at the sample point is above 0.1 ppm. Ecotao supplies it across South Africa and neighbouring countries.

Eco Sensors UV-100 ozone in air analyser with digital ppm display and LED status indicators
  • 0.01–999 ppmRange, in three sub-ranges
  • 2%Accuracy
  • 10 sSampling time
  • 254 nm UVSame principle as the SA reference method

What is ozone in air, and why measure it?

Ozone (O3) is a highly reactive form of oxygen. It is a strong oxidant, which makes it useful for disinfection, odour control and water treatment. The same property makes it a respiratory irritant, so concentrations in occupied spaces have to be kept low and checked.

Where ozone in air comes from

  • Ozone generators (corona discharge and UV types) used for water treatment, food processing, cold-room sanitation and odour removal.
  • Ozone leaks from pipework, contact tanks and off-gas destruct units in treatment plants.
  • Electrical and UV equipment such as high-voltage discharge, some welding processes and 185 nm UV lamps.
  • Photochemical smog, where sunlight drives reactions between nitrogen oxides and volatile organic compounds outdoors.

Why measure it

  • Worker safety: confirm exposure stays below the South African occupational limits (see limits table).
  • Process control: hold a target ozone concentration in a treatment chamber, test cabinet or room-disinfection cycle.
  • Re-entry checks: confirm a room or cold store is safe after ozone treatment.
  • Reference checks: verify the readings of lower-cost electrochemical or metal-oxide ozone monitors.

Units at a glance: 1 ppm = 1,000 ppb. At 25 °C and 101.3 kPa, 1 ppb of ozone ≈ 1.96 µg/m³, so 0.1 ppm ≈ 196 µg/m³ (0.2 mg/m³) and South Africa's ambient standard of 120 µg/m³ equals 61 ppb.

How a UV photometric ozone analyser works

Ozone has a very strong absorption band in the ultraviolet, peaking close to the 253.7 nm emission line of a low-pressure mercury lamp. A UV photometer uses that fact to count ozone molecules directly, rather than inferring them from a chemical reaction.

  1. Sample: air is drawn from the measuring point through tubing and a filter into an optical cell.
  2. Absorb: a 254 nm mercury lamp shines through the cell. Ozone in the sample absorbs part of that UV light.
  3. Detect: a UV detector measures how much light gets through. More ozone means less light.
  4. Calculate: the Beer–Lambert law converts light absorption to concentration. Temperature and pressure in the cell are measured and corrected for, so the result is reported accurately in ppm (v/v).

Because the measurement depends on a physical property of the ozone molecule, UV photometry is stable and largely free of the cross-sensitivity to chlorine, nitrogen dioxide and humidity that affects chemical sensors. This is why South Africa's National Ambient Air Quality Standards specify the UV photometric method (SANS 13964) as the reference method for ozone.

Eco Sensors UV-100 ozone analyser

Eco Sensors UV-100 UV photometric ozone analyser, 0.01 to 999 ppm, front panel with LCD and LEDs
UV photometric · 0.01–999 ppm

UV-100: laboratory-grade ozone measurement in a 2.1 kg instrument

The UV-100 brings the measurement principle used in reference-grade ozone monitors into a benchtop instrument you can carry between sites. It reads ozone in air from 0.01 ppm, low enough to check workplace exposure against the 0.1 ppm limit, up to 999 ppm for ozone treatment chambers and diluted process gas. Readings update every 10 seconds.

Three selectable ranges (0–1.00 ppm, 0–10.0 ppm and 0–999 ppm) let you match the display resolution to the job. Internal temperature and pressure sensors correct the reading automatically, which matters for accurate results at altitude on the Highveld. Each instrument is calibrated against a reference instrument with NIST-traceable calibration, and the calibration data and date are included in its document pack.

  • Built-in data logger downloads over USB or serial as a time- and date-stamped record, for trend analysis and audit trails.
  • 4–20 mA and 0–2.5 VDC outputs connect to a PLC, SCADA system, chart recorder or ozone generator control loop.
  • Four status LEDs show power on, recalibration due, sample filter change due, and ozone above 0.1 ppm at the sample point.
  • Portable or fixed: runs on an AC adapter or a 12 V external battery (12 VDC at 400 mA), so it works in the lab, on a plant walkdown or in a vehicle.
  • Compact: 95 × 210 × 216 mm and 2.1 kg, with a ¼" compression fitting for the sample inlet.

What the UV-100 does in each application

Water and wastewater treatment

Monitors ozone leaking into generator rooms and around contact tanks, and checks that off-gas destruct units are working. The 4–20 mA output can drive an alarm or ventilation interlock.

Room and cold-store disinfection

Confirms the target ozone dose was reached during a treatment cycle, then confirms levels have fallen below 0.1 ppm before staff re-enter. The log provides a record of each cycle.

Food and agriculture

Measures ozone in fruit stores, packhouses and grain fumigation enclosures, where too little ozone fails to control mould and too much damages produce and equipment.

Laboratory and research

Gives accurate, repeatable concentration data for oxidation experiments, material ageing tests and university air chemistry studies, without the drift of chemical sensors.

Checking other ozone monitors

Acts as a reference when bump-testing electrochemical or metal-oxide ozone monitors, showing whether a fixed alarm still reads correctly after months in service.

Occupational hygiene surveys

Maps ozone around photocopier rooms, UV curing lines, welding bays and high-voltage equipment, with data that supports an exposure assessment against the SA limits.

UV-100 specifications

← Scroll to see full table →

SpecificationUV-100
Measurement principleUV absorption photometry
UV source254 nm mercury lamp
Measuring range0.01–999 ppm (v/v)
Selectable ranges0–1.00 ppm, 0–10.0 ppm, 0–999 ppm
Accuracy2%
Sampling time10 seconds
CompensationInternal temperature and pressure correction
OutputsDigital display, 0–2.5 VDC, 4–20 mA, USB / serial data
Data loggingBuilt-in, time- and date-stamped
LED indicatorsOn/off, recalibrate, change filter, >0.1 ppm at sample area
Sample inlet¼" compression fitting
Warm-upAt least 30 minutes; over 1 hour preferred for best accuracy
Power12 VDC at 400 mA (AC adapter or 12 V external battery)
Dimensions95 × 210 × 216 mm (3.75 × 8.25 × 8.50 in)
Weight2.1 kg (4.7 lb)
CalibrationAgainst a reference instrument with NIST-traceable calibration; certificate data supplied
WarrantyOne year

Using it in cold or humid conditions: where the sample air is cold and condensing, such as a cold store, use extended sample tubing and a moisture trap so water doesn't condense in the optical cell.

How to choose an ozone in air instrument

Three sensing technologies are used to measure ozone in air. The right one depends on how accurate the result must be, what concentration you expect, and whether you need a spot reading, a record or an automatic alarm.

← Scroll to see full table →

FactorUV photometric analyser (e.g. UV-100)Electrochemical sensorHeated metal-oxide (HMOS) sensor
How it senses ozoneAbsorption of 254 nm UV light by ozone moleculesChemical reaction at an electrode produces a currentOzone changes the electrical resistance of a heated oxide film
Accuracy and stabilityHighest; no sensor consumptionGood when freshly calibrated; drifts as the cell agesModerate; affected by humidity and temperature
Cross-sensitivityLow; few other gases absorb at 254 nm in normal airResponds to other oxidants such as chlorine and nitrogen dioxideCan respond to humidity changes and some VOCs
Typical roleReference readings, lab work, process control, checking other monitorsHandheld leak detection, personal and fixed safety alarmsLow-cost fixed monitors and controllers
Calibration needsPeriodic check against a referenceRegular bump tests and sensor replacementRegular calibration; sensor head replacement
SizeBenchtop / portable (2.1 kg)Pocket to wall-mountPocket to wall-mount

Decision checklist

  1. Do you need a defensible number? For lab data, process validation, audits or checking other instruments, choose UV photometry.
  2. What concentration will you see? Occupied spaces are usually below 0.2 ppm; treatment chambers can run to tens of ppm. The UV-100's three ranges cover both, from 0.01 to 999 ppm.
  3. Are other oxidising gases present? Chlorine (in water treatment) or nitrogen dioxide (near combustion) can give false readings on electrochemical sensors. UV avoids this.
  4. Do you need control or a record? Choose an instrument with 4–20 mA output for PLC or generator control, and a data logger for audit trails. The UV-100 has both.
  5. Do you need a permanent alarm in every room? Fixed electrochemical or HMOS monitors are more economical for multi-point alarms. Use a UV analyser to check them. See our ozone monitors and controllers.

Ozone exposure limits in South Africa

South Africa regulates ozone in two places: workplace air, under the Occupational Health and Safety Act, and outdoor ambient air, under the National Environmental Management: Air Quality Act.

← Scroll to see full table →

RegulationConditionLimit
Regulations for Hazardous Chemical Agents, 2021 (OEL-RL, 8-hour TWA)Heavy work0.1 ppm
Moderate work0.16 ppm
Light work0.2 ppm
Any workload, 2 hours or less0.4 ppm
National Ambient Air Quality Standards (GN 1210 of 2009)8-hour running average120 µg/m³ (61 ppb); 11 exceedances allowed per calendar year
Reference methodUV photometric method, SANS 13964

The UV-100's sample-area LED lights above 0.1 ppm, the strictest of the workplace limits. Its 4–20 mA output can trigger extract fans or an alarm at a set point you choose.

South African conditions to plan for

Altitude on the Highveld

At Johannesburg and Pretoria altitudes the air is noticeably less dense than at the coast. A UV photometer counts ozone molecules in its cell, so it must correct for pressure to report ppm correctly. The UV-100 does this automatically with internal temperature and pressure sensors. Remember that µg/m³ conversions also shift with local pressure, while ppm (v/v) does not.

Ambient compliance monitoring

The UV-100 uses the same UV photometric principle as the NAAQS reference method and suits indoor, workplace and process measurements. If you are reporting ambient air quality to a licensing authority, confirm what instrument and quality-assurance requirements apply before choosing equipment. Ask Ecotao to talk through your application.

Ozone in air: frequently asked questions

How is ozone in air measured?

The most accurate method is UV photometry. Ozone strongly absorbs ultraviolet light at 254 nm, so an analyser draws air through an optical cell, measures how much 254 nm light is absorbed, and converts that to a concentration using the Beer–Lambert law. South Africa's National Ambient Air Quality Standards name the UV photometric method (SANS 13964) as the reference method for ozone. Lower-cost electrochemical and heated metal-oxide (HMOS) sensors are used for handheld monitors, leak detection and fixed alarms.

What is a safe level of ozone in air in South Africa?

For workplaces, the Regulations for Hazardous Chemical Agents, 2021 set an 8-hour occupational exposure restricted limit for ozone of 0.1 ppm for heavy work, 0.16 ppm for moderate work and 0.2 ppm for light work, and 0.4 ppm for any workload lasting two hours or less. For outdoor ambient air, the National Ambient Air Quality Standards set 120 µg/m³ (61 ppb) as an 8-hour running average, with 11 exceedances allowed per calendar year.

What is the difference between a UV ozone analyser and an electrochemical ozone sensor?

A UV ozone analyser measures the light absorption of ozone itself, so it is accurate, stable and largely free of the cross-sensitivity that affects chemical sensors. It is the right choice for reference measurements, laboratory work and checking other monitors. An electrochemical sensor reacts chemically with oxidising gases, so it is smaller and cheaper but can also respond to gases such as chlorine and nitrogen dioxide, and it drifts as the sensor ages. It suits handheld leak detection and fixed safety alarms.

What range does the UV-100 ozone analyzer measure?

The Eco Sensors UV-100 measures ozone from 0.01 to 999 ppm (v/v) in three selectable ranges: 0–1.00 ppm, 0–10.0 ppm and 0–999 ppm, with 2% accuracy and a 10-second sampling time.

How do I convert ozone ppm to µg/m³?

At 25 °C and 101.3 kPa, 1 ppb of ozone is about 1.96 µg/m³ and 1 ppm is about 1,960 µg/m³ (1.96 mg/m³). This is why South Africa's ambient standard of 120 µg/m³ corresponds to 61 ppb. Because µg/m³ depends on temperature and pressure, readings at altitude, for example on the Highveld, convert slightly differently, while ppm (v/v) is a volume ratio that does not change with pressure.

Does altitude affect ozone readings in Johannesburg and on the Highveld?

Yes, for uncompensated optical instruments. A UV photometer measures the number of ozone molecules in its cell, and at Highveld altitude the air is less dense, so the instrument must correct for pressure and temperature to report ppm accurately. The UV-100 has internal temperature and pressure sensors and corrects its calibration automatically.

How long does the UV-100 need to warm up?

The published operating instructions call for at least 30 minutes of warm-up, with more than one hour preferred for best accuracy, because the UV lamp and optical bench need to stabilise.

Can the UV-100 log data or connect to a PLC?

Yes. It has a built-in data logger that can be downloaded over USB or serial as a time- and date-stamped data stream, plus 0–2.5 VDC and 4–20 mA analogue outputs for chart recorders, PLCs, SCADA systems or ozone generator control loops.

Where can I buy an ozone in air analyser in South Africa?

Ecotao supplies the Eco Sensors UV-100 ozone analyzer from Cape Town, with delivery across South Africa and to Namibia, Botswana, Zimbabwe, Mozambique, Lesotho and Eswatini. Contact sales@ecotao.co.za or +27 21 911 5835, or check stock on the Ecotao store.

Need help choosing an ozone in air instrument?

Tell us where you're measuring, the concentrations you expect and whether you need logging or control outputs. We supply across South Africa, Namibia, Botswana, Zimbabwe, Mozambique, Lesotho and Eswatini.