PC 50 & PC 60

PC 50

pH – mV – ORP – Cond – TDS – Temperature

Professional bench meters with innovative backlit high defi nition
colored LCD display, with possibility of brightness adjustment.
Management of DHS digital sensors.
Possibility to select the parameters to show on the display.
The step by step guide in setup and calibration allows a simple and immediate use of the device.
Automatic pH calibration with USA and NIST buffers, up to 3 points and 2 points defi ned by the user.
Automatic ORP calibration on 1 point.
Automatic conductivity calibration up to 5 points (84; 147; 1413 μS e 12,88; 111,8 mS) and 1 point custom.
Automatic resolution, reference temperature from 15 to 30 °C, and selectable cell costant (C= 0,1; 1; 10).
Indication of the calibrated points, with icons.
Measurement stability indication and possibility to select 3 stability levels.
All operations are constantly monitored and shown to the operator by the colored LED above the display and
self-diagnosis messages.

PC 60

pH – mV – ORP -Cond – TDS – Salinity – Temperature – GLP – Data Logger

… only on PC 60 VioLab
GLP function: Date and time, recall of calibration data
and function “Calibration due”.
Data logger up to 1000 savings, manual and automatic, PC connect via micro-USB.
Data-Link + software available from www.xsinstruments.com.
RS 232 external printer connectio